Welcome to the official website of Ningbo Kesheng Experimental Instrument Co., Ltd., service hotline: 0574-88815882
News Center
Engaged in software and hardware development of scientific research instruments
顶部菜单
The Use and Daily Cleaning of Carbon Dioxide Incubators
Author:管理员    Dtime:2021-03-24 10:56:07    Browse (2919)
Memo:A carbon dioxide incubator is a device that simulates the formation of a growth environment similar to that of cells/tissues in an organism. The incubator requires a stable temperature (37 ° C), a stable CO2 level (5%), a constant pH (7.2-7.4)......

A carbon dioxide incubator is a device that simulates the formation of a growth environment similar to that of cells/tissues in an organism. The incubator requires a stable temperature (37 ° C), a stable CO2 level (5%), a constant pH (7.2-7.4), and a high relative saturation humidity (95%) to culture cells/tissues in vitro.
1、 Precautions for use
1. Do not turn on the power switch before filling the carbon dioxide incubator with water, otherwise it will damage the heating element.
2. After running the incubator for several months, the water in the tank may decrease due to volatilization. When the low water level indicator light (WLow12) is on, water should be added. First, open the overflow pipe and use a funnel to connect a rubber tube to add water from the water injection hole to turn off the low water level indicator light. Then, measure and add water (CP-ST200A adds 1800ml of water, CP-ST100A adds 1200ml of water), and then block the overflow hole.
3. The carbon dioxide incubator can be used as a high-precision constant temperature incubator, and the CO2 control system must be turned off at this time.
4. Because the CO2 sensor is calibrated under saturated humidity, the humidification plate must always be filled with sterilized water.
5. When the displayed temperature exceeds the set temperature by 1 ℃, the over temperature alarm indicator light (OverTemp7) will light up and emit a sharp alarm sound. At this time, the power should be turned off for 30 minutes; If the power (temperature control) switch (Power20) is still overheated when turned on again, the power should be turned off and the maintenance personnel should be notified. (Test article)

搜索

Copyright: Ningbo Kesheng Experimental Instrument Co., Ltd        浙ICP备18036608号-1  

Production Base: No. 90 Jindi Temple, Shangtian Street, Fenghua District, Ningbo City, Zhejiang Province

E-mail:ningbokesheng@163.com

Technical QQ:2742476083       Sales QQ:2903595711

Contact information

Contact us

Website Navigation

 
 

HOME         ABOUT          PRODUCTS

NEWS     APPLICATION     CONTACT

Official website mobile end
WeChat contact person

Service Hotline :0574-8881 5882

Sales department switchboard:0574-8791 6861

Customer Service:180 6726 8163

Technical support:189 5821 5885

After-sale service:181 0012 1810

Customer
service

Online Service Service time :9:00-18:00

Select the following customer service representatives:

Service
hotline

0574-88815882
Customer service consultation hotline

Learn
more

Scan code to add